ABSS calibration
Corrections for Swisstrace ABSS are done in PMOD PSAMPLE software. This page applies to the other blood sampling systems.
Calibration of ABSS data
Calibration of automatic blood sampling systems data includes
- ABSS specific detector-to-gamma counter coefficient
- gamma counter-to-pet coefficient, common to all ABSSs
- isotope-specific branching ratio correction
All calibration steps can be done using program blo2kbq (absscal). In short, counts/sec values from ABSS are multiplied by detector calibration factor and gamma counter-to-pet coefficient, and divided by the branching ratio.
Calibration file
In TPC, physicists are responsible of the calibration measurements and the updating of
the calibration file, currently in S:\Lab\plasma\bsampler_calibration\pump_cal.dat
.
The calibration file is in ASCII text format, and it contains the calibration dates, and the above mentioned calibration coefficients.
The contents of the ABSS calibration file on Spring 2014 was:
# YYYY-MM-DD Pump1(ecat) Pump2(ge) Pump3(HR+) Pump4(HRRT) WizardGammacounter # Pump1(PET/CT) 1900-01-01 51 47 1 1 0.00270 1997-07-01 47 42 1 1 0.00270 1999-12-15 43 40 1 1 0.00270 2000-01-13 46.5 42.5 1 1 0.00270 2000-09-06 45.9 44.6 1 1 0.00288 2000-11-15 44.9 42.5 1 1 0.00288 2000-12-05 48.5 45.4 1 1 0.00288 2001-02-02 46.0 41.6 1 1 0.00288 2001-03-28 47.8 44.3 1 1 0.00288 2001-06-07 48.1 43.5 1 1 0.00288 2001-09-18 42 37 1 1 0.00288 2001-11-14 44.5 41.4 1 1 0.00288 2001-12-13 47.5 40 1 1 0.00277 2002-02-07 47.5 144 1 1 0.00277 2002-02-26 45.7 133 1 1 0.00277 2002-03-12 45 40 1 1 0.00277 2002-12-12 45 41 1 1 0.00277 2003-06-07 45 42 1 1 0.00277 2003-08-20 44 41 1 1 0.00278 2004-01-16 47 41 1 1 0.00278 2004-09-09 47 41 14 1 0.00278 2004-09-29 45 40 14 1 0.00278 2005-08-15 45 40 14 1 0.00287 2006-01-20 45 42 14 1 0.00282 2006-04-03 46.5 43 14 1 0.00291 2006-11-01 48.8 43 14 1 0.00291 2006-11-02 48.8 44 14 1 0.00291 2007-10-12 46.5 42.6 13.1 1 0.00283 2008-01-17 46.5 42.6 13.1 1 0.00271 2008-02-14 46.5 42.6 13.1 1 0.00270 2008-03-11 46.8 42.9 13.3 1 0.00270 2008-08-21 46.8 42.9 13.3 1 0.00291 2008-08-22 44.6 40.8 13.5 1 0.00291 2008-09-05 44.6 40.8 13.5 1 0.00283 2008-09-25 44.6 40.8 13.5 1 0.00281 2008-10-30 45.4 41.2 12.6 1 0.00281 2009-01-13 45.6 42.4 13.5 1 0.00281 2009-02-26 47.8 42.9 13.9 1 0.00281 2009-04-02 48.2 43.9 13.6 1 0.00281 2009-04-30 48.2 41.6 13.9 1 0.00281 2009-05-28 46.6 43.1 13.3 1 0.00281 2009-08-24 46.6 43.1 13.3 1 0.00279 2009-12-01 46.6 43.1 13.5 1 0.00279 2010-01-14 46.6 40.6 13.1 12.7 0.00279 2010-01-20 46.6 40.6 13.1 12.7 0.00278 2010-08-11 46.6 40.9 13.3 13.1 0.00278 2010-08-23 46.6 40.9 13.3 13.1 0.00282 2010-09-23 46.6 40.9 13.2 12.5 0.00282 2011-08-12 46.6 40.9 13.6 12.5 0.00282 2011-08-23 46.6 40.9 13.6 12.5 0.00279 2012-08-15 46.6 40.9 13.6 12.5 0.00288 2013-01-15 46.6 40.9 13.1 12.7 0.00274 2013-05-03 46.6 45.0 13.1 12.7 0.00274 2013-09-11 46.6 45.0 13.1 13.1 0.00274 2014-01-14 46.6 45.0 12.7 12.4 0.00274
See also:
- Processing ABSS data in TPC
- Scanditronics ABSS raw data format
- Allogg ABSS #1 raw data format
- Allogg ABSS #2 raw data format
- Preprocessing arterial input data
Tags: ABSS, Blood, Calibration
Updated at: 2019-12-03
Created at: 2010-09-14
Written by: Vesa Oikonen, Tuula Tolvanen